Thermocouple Wire and Extension Wire Calibration Tolerances
Thermocouple Grade Wire
Calibration Type |
Temperature Range | Tolerance Whichever is GREATER | ||
---|---|---|---|---|
J | 32 to 1382°F | 0 to 750°C | ± 2.2°C or ± 0.75% | ± 1.1°C or ± 0.4% |
K | 32 to 2282°F | 0 to 1250°C | ± 2.2°C or ± 0.75% | ± 1.1°C or ± 0.4% |
E | 32 to 1652°F | 0 to 900°C | ± 1.7°C or ± 0.5% | ± 1.0°C or ± o.4% |
T | 32 to 662°F | 0 to 350°C | ± 1.0°C or ± 0.75% | ± 0.5°C or ± 0.4% |
R or S | 32 to 2642°F | 0 to 1450°C | ± 1.5° or ± 0.25% | ± 0.6°C or ± 0.4% |
B | 1598 to 3092°F |
870 to 1700°C | ± 0.5% | |
N | 32 to 2282°F | 0 to 1250°C |
± 2.2°C or ± 0.75% | ± 1.1°C or ± 0.4% |
Thermocouple Extension Grade Wire
Calibration Type |
Temperature Range | Tolerance Whichever is GREATER | ||
---|---|---|---|---|
JX | 32 to 392ºF | 0 to 200°C | ± 2.2°C | ± 1.1°C |
KX | 32 to 392°F | 0 to 200°C | ± 2.2°C | ± 1.1°C |
EX | 32 to 392°F | 0 to 200°C | ± 1.7°C | ± 1.0°C |
TX | 32 to 212°F | 0 to 200°C | ± 1.0°C | ± 0.5°C |
NX | 32 to 392°F | 0 to 200°C | ± 2.2° | ± 1.1°C |
Compensating Extension Cable
Calibration Type |
Temperature Range | Tolerance | |
---|---|---|---|
BX | 32 to 392ºF | 0 to 200°C | ± 3.7°C |
CX | 32 to 1600°F | 0 to 870°C | ± 6.8°C |
RX or SX | 32 to 392°F | 0 to 200°C | ± 5.0°C |
Reference junction 32ºF (0ºC)
Search Our Site
Other Thermocouple Wires
- PVC 220°F 105°C
- Synthetic Fiber 550°F 290°C
- PVC/Shielded/PVC 220°F 105°C
- PVC/Individual and Overall Shield/PVC 220°F 105°C
- FEP 400°F 205°C
- FEP/Shielded/FEP 400°F 205°C
- PFA 500°F 260°C
- PFA/Shielded/PFA 500°F 260°C
- TFE 500°F 260°C
- TFE/Shielded/TFE 500°F 260°C
- Kapton 500°F 260°C
- Kapton/Shielded/Kapton 500°F 260°C
- Fiberglass 900°F 482°C
- High Temperature Fiberglass 1200°F 650°C
- Refrasil 1600°F 871°C
- High Temperature “S” Glass 1200°F 650°C
- Ceramic Fiber 2200°F 1205°C
- Mica Ceramic Fiber 2200°F 1205°C
- Metal Overbraid
- International Thermocouple and Extension Wire Color Codes
- Thermocouple Wire and Extension Wire Calibration Tolerances